PCB-3713B11200G
- Triaxial MEMS accelerometer,
- Suitable for measuring static accelerations,
- Measuring range 200 g
- Sensitivity 10 mV / g,
- Frequency range 0 … 850 Hz,
- Supply voltage 6 … 30 V DC,
- 9-pin female connector on the side
basic data | |||
Sensitivity (± 10%) | 10 mV / g | 1.0 mV / (m / s) | [1] |
measuring range | ± 200 g pk | ± 1962.0 m / s pk | |
Frequency range (± 5%) | 0 ... 850 Hz | ||
Frequency range (± 10%) | 0 ... 1250 Hz | ||
resonant frequency | ?11.000 Hz | ||
Phase response (100 Hz) | <10 ° | ||
Broadband resolution | 21.1 mg rms | 0.21 m / s RMS | [2] |
linearity error | ≤1% | [3] | |
Lateral acceleration sensitivity | ?3% | ||
Operating conditions | |||
Overload limit (shock) | ± 3.000 g pk | ± 29,420 m / s² pk | |
The temperature range (operation) | -65 ... 250 ° F | -54 ... 121 ° C | |
Temperature range (storage) | -65 ... 250 ° F | -54 ... 121 ° C | |
Temperature coefficient of sensitivity | ± 3% | [4] | |
Temperature coefficient of the zero point | ± 2% FSO | [4] [5] | |
Strain sensitivity (base surface) | 0.0005 g / µε | 0,005 (m / s²) / µε | [2] |
magnetic sensitivity | 70 ug / µε | 6.9 (m / s²) / T | [2] |
Technical characteristics | |||
supply voltage | 6 ... 30 V DC | ||
current consumption | ?6 mA | ||
output impedance | ≤100 Ω | ||
voltage offset | ± 40 mV DC | [6] | |
Spectral Noise (1) | 2518 µg/√Hz | 24,693 (µm/s²)/√Hz | [2] |
Electrical insulation (housing) | > 100 M | ||
physical properties | |||
housing material | titanium | ||
seal | hermetically sealed | ||
Size - Length | 0.8 in | 20.3 mm | |
Size - Width | 0.8 in | 20.3 mm | |
Size - Height | 0.8 in | 20.3 mm | |
Dimensions | 0.6 oz | 17.3 grams | |
electrical connection | 9-Pin | ||
port location | Laterally | ||
mounting thread | 10-32 female thread | ||
Hints | |||
[1] | Measured at 100 Hz, 10 g RMS | ||
[2] | typically, | ||
[3] | Zero-based line of best fit, determined by the method of least squares | ||
[4] | -54 ... 121 ° C, Reference 24 ° C | ||
[5] | FSO = "Full Scale Output": maximum output signal over the measuring area (4 V DC ). | ||
[6] | Offset at 3 meter cable (type 037) | ||
[7] | Details: see PCB Declaration of Conformance PS027 | ||
included supply | |||
080A12 | adhesive mounting | -1 | |
081B05 | Mounting bolts (10-32 to 10-32) | -1 | |
ACS 103T | Calibration of the amplitude and phase response of 2 Hz to the upper 5% -frequency for triaxial transducer | -1 | |
M081B05 | Mounting bolts 10-32 on M6 x 0.75 | -1 |